Description
AD8000 is a microprocessor-based pH, ORP, conductivity, TDS and temperature benchtop meter. Relative mV feature is also provided. The auto ranging feature of the EC and TDS ranges automatically sets the instrument to the scale with the highest resolution. pH measurements are compensated for temperature effect automatically (ATC) using AD7662 temperature probe, while the EC readings can be compensated automatically (ATC) using the conductivity probe with a built-in temperature sensor or manually (MTC). It is also possible to disable the temperature compensation and measure the actual conductivity. The temperature coefficient is user selectable. The pH calibration can be performed at 1, 2 or 3 points with five memorized buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01). In addition, AD8000 is provided with calibration timeout alarm, GLP feature to view last calibration data and automatic HOLD feature to freeze the first stable reading on the LCD.
pH Specifications
pH range
-2.00 to 16.00 pH / -2.000 to 16.000 pH
pH Resolution
0.01 pH / 0.001 pH
pH Accuracy (@25ºC/77ºF)
±0.01 pH / ±0.002 pH
pH Calibration
Up to 3-point calibration, 5 buffer available (4.01, 6.86, 7.01, 9.18, 10.01 pH)
ORP Specifications
ORP range
±2000 mV
ORP Resolution
±699.9 mV / ±2000 mV
ORP Accuracy (@25ºC/77ºF)
0.1 mV (±699.9 mV) / 1 mV (±2000 mV)
ORP Calibration
Factory calibrated
EC Specifications
EC range
0.00 to 19.99 μS/cm | 20.0 to 199.9 μS/ cm | 200 to 1999 μS/cm | 2.00 to 19.99 mS/cm | 20.0 to 199.9 mS/cm
EC Resolution
0.01 μS/cm | 0.1 μS/cm | 1 μS/cm | 0.01 mS/cm | 0.1 mS/cm
EC Accuracy (@25ºC/77ºF)
±1% f.s.
EC Calibration
Offset at 0.00 μS/cm; Slope at 1 point with 6 memorized values (84.0, 1413 μS/cm; 5.00, 12.88, 80.0, 111.8 mS/cm) or with custom value
TDS Specifications
TDS range
0.00 to 9.99 ppm | 10.0 to 99.9 ppm | 100 to 999 ppm | 1.00 to 9.99 ppt | 10.0 to 99.9 ppt
TDS Resolution
0.01 ppm | 0.1 ppm | 1 ppm | 0.01 ppt | 0.1 ppt
TDS Accuracy (@25ºC/77ºF)
±1% f.s.
Temperature Specifications
Temperature Range
-9.9 to 120 °C (pH range) | 0.0 to 100.0°C (EC range)
Temperature Resolution
0.1°C
Temperature Accuracy
±0.5°C
Additional Specifications
pH Electrode/Probe
AD1131B - Refillable, glass body pH electrode with double junction, BNC connector and 1 m cable (Included)
Temperature Electrode/Probe
AD7662 - Stainless steel temperature probe with 1 m cable (Included)
GLP
View last calibration data
Hold function
Yes
Stability indicator
Yes
Low battery indication
Yes
Power supply
12 Vdc power adapter (included)
Weight
1.8 kg
Dimensions
230 x 180 x 50 mm
Ordering Information
Adwa AD8000 is supplied with AD1131B - Refillable glass body pH electrode with BNC connector and 1 m cable; AD76309 - Conductivity probe with built-in temperature sensor; AD7662 - Stainless steel temperature probe with 1 m cable; Electrode holder; AD70004 - pH4 calibration solution in sachet (20 ml); AD70007 - pH7 calibration solutions in sachet (20 ml); AD70031 - 1413 μS/cm calibration solutions in sachet (20 ml); AD70030 - 12.88 mS/cm calibration solutions in sachet (20 ml); 12 Vdc power adapter; Instruction manual.