ADWA AD310 Standard Professional Conductivity-Temperature Portable Meter

0.00

Auto-Ranging feature that sets the instrument to the scale with the highest resolution; Measurement stability indicator

  • Parameters: EC(Conductivity), Temperature
Compare

Description

AD310 is a portable microprocessor- based instrument for measuring EC, and temperature. The auto ranging feature of the EC readings automatically sets the instrument to the scale with the highest resolution. Measurements are compensated for temperature effect automatically (ATC) or manually (MTC). The temperature compensation feature can also be disabled. The temperature coefficient is user selectable. The instrument also features a measurement stability indicator.

EC Specifications

EC range

19.99 μS/cm | 199.9 μS/cm | 1999 μS/cm | 19.99 mS/cm | 199.9 mS/cm

EC Resolution

0.01 μS/cm | 0.1 μS/cm | 1 μS/cm | 0.01 mS/cm | 0.1 mS/cm

EC Accuracy (@25ºC/77ºF)

±1% f.s.

EC Calibration

Offset: 0.0 μS/cm; Slope: 1 point with 6 memorized standards (84.0, 1413 μS/cm; 5.00, 12.88, 80.0, 111.8 mS/cm) or one custom value

Temperature Compensation

ATC/MTC, -9.9 to 120.0°C

Temperature Specifications

Temperature Range

-9.9 to 120.0°C

Temperature Resolution

0.1°C

Temperature Accuracy

±0.5°C

Additional Specifications

Electrode/Probe

AD76309 - Stainless steel, 4-ring EC/TDS/Temperature probe with DIN connector and 1 m cable

Hold Function

Yes

Low Battery Indication

Yes

Auto Shutt-off

After 8 minutes of non-use conserves energy and lengthens battery life-span

Battery Type / Life

4 x 1.5V AA batteries / Approx. 1500 hours of continuous use

Dimensions

70 x 145 x 26 mm

Weight

Approx. 200 g

Ordering Information

Adwa AD310 is supplied with AD76309 - EC/TDS/Temperature probe with builtin temperature sensor and 1 m cable; AD70031 - 1413 μS/cm calibration solutions in sachet (20 ml); Batteries (4 x 1.5V, AA alkaline); Instruction manual